COMPLIANCE

Range of frequencies (20 Hz to 1 MHz) and test signal levels (5 mV to 2 Vrms, 50 mA to 20 mArms). With Option 001 the HP 4284A's test signal level range spans 5 mV to 20 Vrms, and 50 mA to 100 mArms.


MEASUREMENT FUNCTIONS

TEST SIGNAL

Frequency:
20 Hz to 1 MHz, 8610 selectable frequencies
(refer to APPENDIX G for selectable frequencies)

Accuracy:
+/- 0.01 %

Signal Modes:
        Normal:    Programs selected voltage or current at the measurement terminals when they are opened or shorted, respectively.
        Constant:  Maintains selected voltage or current at the device under test independent of changes in the device's impedance.

Signal Level:
 
Mode
Range
Setting Accuracy
    Voltage 
    Non-constant 
Constant *
5 mVrms to 2 Vrms
10 mVrms to 1 Vrms
+/- (10% + 1 mVrms)
+/- (6% + 1 mVrms)
Current
Non-constant
Constant *
50 mArms to 20 mArms
100 mArms to 10 mArms
+/- (10% + 10 mArms)
+/- (6% + 10 mArms)
    * : Automatic Level Control Function is set to ON.

Output Impedance:
100 W +/- 3%

Test Signal Level Monitor:
     Mode 
Range
Accuracy
Voltage 1
5 mVrms to 2 Vrms
0.01 mVrms to 5 mVrms
+/- (3% of reading + 0.5 mVrms)
+/- (11% of reading + 0.1 mVrms)
Current 2
50 mArms to 20 mArms
   0.001 mArms to 50 mArms 
+/- (3% of reading + 5 mArms)
+/- (11% of reading+1 mArms)