DISPLAY PAGES
The 4145b displays six different screens. Each screen is called a
page and each has a different purpose in relation to instrument operation.
By changing from one page to another, different functions and capabilities,
such as measurement set up, measurement, etc., are made available.
Display paging is controlled by the PAGE CONTROL keys
Paging cannot be performed during measurement.
MENU PAGE
The start-up menu is displayed when the instrument is turned on and
each time the MENU key is pressed. Displayed on then MENU page are the
accessible pages, corresponding softkey prompts (SKP).
All pages except the GRAPHICS and LIST pages can be displayed directly
from the MENU page by pressing the appropriate softkey. For example, pressing
softkey 3 while the MENU page is displayed automatically displays the MEAS/DISP
MODE SETUP page.
CHANNEL DEFINITION PAGE
The CHANNEL DEFINITION page is displayed when the softkey 1 or the
NEXT key is pressed while the MENU page is displayed. On this page the
operator must define the name, mode, and function of each channel that
is to be used in the measurement. User functions are also defined on this
page.
SOURCE SETUP PAGE
The SOURCE SETUP page is displayed when softkey 2 is pressed on the
MENU page or when the NEXT key is pressed on the CHANNEL DEFINITION page.
On this page the operator must enter the output parameters (START, STOP,
STEP, COMPLIANCE, etc.) for the source channels (SMUs an voltage sources)
defined on the CHANNEL DEFINITION page.
MEAS/DISP MODE SETUP PAGE
The MEAS/DISP MODE SETUP page is displayed when softkey 3 is pressed
on the MENU page or when the NEXT key is pressed on the SOURCE SETUP page.
On this page the operator must select the desired display mode (GRAPHICS,
LIST), enter the appropriate source and monitor names, and enter the desired
scaling factors.
GRAPHICS PLOT PAGE
The GRAPHICS PLOT page is displayed when the MEAS/DISP MODE SETUP page
has been set up for GRAPHICS MODE DISPLAY and the NEXT key is pressed.
On this page the operator can make the measurement by pressing the SINGLE,
REPEAT, or APPEND key an can analyze the measurement results with the softkeys.
LIST DISPLAY PAGE
The LIST DISPLAY page is displayed when the MEAS/DISP MODE SETUP page
has been set up for LIST mode display and the NEXT key is pressed. On this
page the operator can make then measurement by pressing the SINGLE, REPEAT,
or APPEND key and can analyze the measurement results with the softkeys.
|